Class I and Class II: Two Ways to Be Safe, Two Different Test Sets

Class I equipment relies on basic insulation plus a protective earth connection, so that a fault that breaches the basic insulation becomes a fault current rather than a shock. Class II equipment relies on double or reinforced insulation and has no protective earth connection at all. Both are legitimate; the evidence for each is different.

Class I evidence

Dielectric strength and clearance for the basic insulation, plus the bonding test on every accessible conductive part, plus touch current under normal and single fault conditions. The bonding test is the distinctive one: without it the class I claim has no meaning, because the earth path is the protection.

Class II evidence

Dielectric strength and clearances for reinforced insulation between live parts and every accessible surface, including surfaces that are not obviously conductive, such as a metallised label or a conductive coating. The distinctive test is the assessment of the whole accessible surface, because there is no earth to fall back on.

The mixed construction

Many instruments are Class I with Class II subassemblies, or Class II with a functional earth. The rules then apply separately to each part and to the interfaces between them. A report that assigns a single class to a mixed construction is under-specified, and the ambiguity tends to appear at the exact interface where the insulation changes.

Related technical notes

    Standard:       IEC 61032 IEC 60061 IEC 60335 IEC 60529 IEC 60068 IEC 60950 IEC 61010 IEC62368 IEC 60601
    TAG:
Test Probe Go Not Go Gauges Spring Impact Hammer Test Pin Probe Test Probe B Test Probe 18 IP Test Probe
 
Lampholder Gauges Lampcap Gauge Test Probe 11 UL Test Finger Probe Test Vessel Rigid Test Finger Impact Testing
test probe