Temperature rise testing checks that no part of the equipment becomes hot enough to be a burn hazard, to damage insulation, or to ignite surrounding material. The limits are stated as rises above ambient rather than absolute temperatures, because the equipment is rated for an ambient range and must remain safe at the top of it.

The ambient assumption

Unless the manufacturer declares otherwise, the reference ambient is conventionally 40 °C for this family of equipment. A measurement taken in a 22 °C laboratory therefore has to be compared with the limit using the rated ambient, not the laboratory temperature. Reporting a 30 K rise as acceptable because the part measured 52 °C in a cool room is the single most common arithmetic error in thermal testing.

Which parts

Windings, transformer cores, wiring insulation, terminals, the enclosure surface, and any part the user can touch. The last one has the lowest limit and is the one most often skipped, because the thermocouple is placed inside the case where the interesting temperatures are rather than on the handle where the user is.

Test conditions

Worst case means the highest rated supply voltage, the maximum rated load, and the least favourable mounting. A bench instrument tested in free air on an open bench will run cooler than the same instrument in a rack, and the rack is a legitimate installation the instructions usually permit.

Related technical notes

    Standard:       IEC 61032 IEC 60061 IEC 60335 IEC 60529 IEC 60068 IEC 60950 IEC 61010 IEC62368 IEC 60601
    TAG:
Test Probe Go Not Go Gauges Spring Impact Hammer Test Pin Probe Test Probe B Test Probe 18 IP Test Probe
 
Lampholder Gauges Lampcap Gauge Test Probe 11 UL Test Finger Probe Test Vessel Rigid Test Finger Impact Testing
test probe