Battery-powered instruments are in scope of IEC 61010 and the battery introduces risks that the mains supply does not: a high short circuit current at low voltage, venting of gas during overcharge, leakage of electrolyte, and the possibility of the user fitting the battery the wrong way round.
Charging
Where the instrument charges a battery internally, the charging circuit is assessed for overcharge protection, temperature monitoring where the chemistry requires it, and separation between the mains side and the battery side. A charger that relies on the battery protection circuit alone is relying on a component that may be absent in a replacement battery.
Venting and leakage
The enclosure must tolerate the gas a venting cell produces without creating an explosive atmosphere or a flame path to the outside, and must contain leaked electrolyte away from live parts and from the user. Sealed lead acid and lithium chemistries both need this assessment, with different failure modes.
Reverse polarity and user replacement
If the instructions permit the user to replace the battery, the equipment must be safe with the battery inserted either way, or the battery compartment must make reversal physically impossible. Assuming the user reads the polarity marking is not an accepted measure, because the standard is written for the case where the user does not.