Protective Bonding: The 0.1 Ohm Test and What It Really Checks

On Class I equipment, the protective bond connects every accessible conductive part to the protective earth terminal. The test injects a defined current, typically 25 A from a low voltage source, and requires that the voltage drop corresponds to a resistance below a small limit, conventionally 0.1 ohm.

Why a high current

A low current measurement through a multimeter reads the contact resistance of the probes as much as the bond. A 25 A test is high enough that a poor bond, such as a painted surface under a screw or a single strand of a stranded wire, heats and shows a resistance far above the limit. The test is designed to find the connections that would fail to carry fault current, not to measure a precise value.

Where bonds fail

Under paint and anodising, at screw threads where a star washer was omitted, at hinges and sliding joints that were never bonded at all, and at cable screens terminated by a drain wire that does not make contact. All four are construction issues that a drawing review will not catch and a single measurement will.

Doing it correctly

Measure from the earth pin of the plug or the earth terminal to the accessible part, with the probe tip on bare metal rather than on a coating, and with the test current flowing for the specified time. Null the lead resistance before the measurement, and record the value rather than only the pass verdict.

Related technical notes

    Standard:       IEC 61032 IEC 60061 IEC 60335 IEC 60529 IEC 60068 IEC 60950 IEC 61010 IEC62368 IEC 60601
    TAG:
Test Probe Go Not Go Gauges Spring Impact Hammer Test Pin Probe Test Probe B Test Probe 18 IP Test Probe
 
Lampholder Gauges Lampcap Gauge Test Probe 11 UL Test Finger Probe Test Vessel Rigid Test Finger Impact Testing
test probe