Instruments that generate high voltage for testing are a specific case in IEC 61010 because the hazardous voltage is a normal, intended output rather than a fault. The requirements address the internal source, the output terminals, the enclosure, and the behaviour when the output is interrupted.
Internal separation
The high voltage circuit must be separated from the control circuit and from accessible surfaces by insulation appropriate to the working voltage, and the separation must be maintained across the PCB, through the wiring harness, and at the transformer. Wiring that crosses the barrier without a defined route, or a harness that can move against a sharp edge, is a common finding.
Interlocks and warning
Where the enclosure can be opened, an interlock that removes the high voltage, plus a warning that the output may remain charged, is the expected pattern. A warning label alone is not an interlock, and an interlock that can be held closed with tape is not a measure.
Discharge
Capacitance in the output circuit stores energy after the test ends. The design must discharge it within a defined time to a defined level, or the terminals must be marked as retaining charge and the instructions must require a discharge procedure. Measuring the decay with a high impedance voltmeter is part of the test, not an optional extra.