Altitude Correction: Why Clearances Grow Above 2000 m

Clearance requirements in IEC 61010 assume operation at altitudes up to 2000 m. Above that, the reduced air pressure lowers the breakdown voltage of a given gap, and the standard requires either increased clearances or a reduction in the working voltage that the equipment is rated for.

The physics

Breakdown voltage depends on gas density, which falls with altitude at constant temperature. At 3000 m the density is roughly 70 percent of sea level value, and at 5000 m it is around 55 percent. A gap that is comfortable at sea level can flash over at altitude, and the failure is a function of the transient overvoltage the installation can deliver, not of the normal working voltage.

How the standard handles it

Either the manufacturer declares an altitude range and applies a correction factor to the clearance table, or the equipment is rated only for 2000 m and the instructions state that. Declaring a high altitude without changing the design is a marking claim with no supporting evidence, and it is the kind of claim that fails when equipment is installed in a mountain facility.

Practical responses

Increase clearances, pot or encapsulate the critical gaps, or reduce the rated working voltage at altitude and say so in the manual. The third option is often the cheapest and is rarely chosen, because a reduced rating is less attractive on a datasheet than an unverified altitude claim.

Related technical notes

    Standard:       IEC 61032 IEC 60061 IEC 60335 IEC 60529 IEC 60068 IEC 60950 IEC 61010 IEC62368 IEC 60601
    TAG:
Test Probe Go Not Go Gauges Spring Impact Hammer Test Pin Probe Test Probe B Test Probe 18 IP Test Probe
 
Lampholder Gauges Lampcap Gauge Test Probe 11 UL Test Finger Probe Test Vessel Rigid Test Finger Impact Testing
test probe