The Dielectric Strength Test: Setup Details That Decide the Result

Dielectric strength testing applies a high voltage between an accessible conductive part and the live parts, or between primary and secondary circuits, and requires that no breakdown occurs. The clause is short; the execution has more variables than the clause suggests.

Voltage and duration

The test voltage depends on the working voltage and on whether the insulation is basic, supplementary or reinforced, and it is applied for a defined period, conventionally one minute. Some standards permit a shorter test at a higher voltage for production testing, and the two are not interchangeable: a routine test is not evidence of type compliance.

Ramp and trip current

The voltage should be raised steadily to the test value rather than switched on at full value, because a step can cause a transient that trips the set and looks like a failure. The trip threshold of the test set matters as much: a set that trips at a few milliamps will fail equipment with legitimate leakage through a filter capacitor, and a set with a very high threshold will pass a genuine partial breakdown.

What counts as failure

Breakdown is a sudden collapse of voltage with a rise in current. A slow increase in current without collapse is usually capacitive charging or corona, and it is not a failure. Distinguishing the two requires watching the current throughout the hold rather than only reading the pass lamp at the end.

Related technical notes

    Standard:       IEC 61032 IEC 60061 IEC 60335 IEC 60529 IEC 60068 IEC 60950 IEC 61010 IEC62368 IEC 60601
    TAG:
Test Probe Go Not Go Gauges Spring Impact Hammer Test Pin Probe Test Probe B Test Probe 18 IP Test Probe
 
Lampholder Gauges Lampcap Gauge Test Probe 11 UL Test Finger Probe Test Vessel Rigid Test Finger Impact Testing
test probe