Two numbers describe the voltage a circuit sees: the nominal voltage of the supply it is connected to, and the working voltage across the insulation being assessed. They are often the same, and when they are not, using the wrong one under-sizes every clearance in the design.
The two definitions
Nominal voltage is the supply the product is designed for: 230 V, 400 V, 120 V. It is a property of the installation.
Working voltage is the highest voltage that can appear across the specific insulation under assessment, in normal operation, including the effects of transformation and resonance.
On a mains input terminal, the two are the same. On the secondary of a transformer, or across the insulation between a switching node and a heatsink, they are not — and the working voltage can be far higher than the supply.
| Location | Nominal | Working voltage |
|---|---|---|
| mains input to earth | 230 V | 230 V |
| transformer secondary | 230 V | whatever the secondary produces |
| switching node to heatsink | 230 V | peak of the switched waveform |
Why the working voltage can exceed the supply
A flyback converter generates a reflected voltage on the primary that adds to the supply. A resonant circuit can ring to twice the supply or more. A capacitive divider can hold a node at a potential that no terminal of the supply provides.
Clearance is set by the voltage that can actually appear, not by the label on the product. A design that applies the mains clearance table to a 600 V switching node has under-sized that gap by a factor of two in distance.
Peak, RMS and the transient
Working voltage is normally expressed as an RMS value for sinusoidal quantities and as a peak for non-sinusoidal ones, and the clearance tables are indexed on the appropriate one.
Mixing them is a common error: taking the RMS value of a waveform with a high crest factor understates the peak by a large margin. A 300 V square wave and a 300 V sine have the same RMS and very different peaks.
Where designs get caught
Isolated outputs. The secondary is treated as low voltage, so the clearance to the primary is designed from the low-voltage table. The insulation between them must withstand the working voltage of the primary side, and the transient that couples across the transformer.
Relays and contactors. The open contact gap sees the supply voltage plus the inductive kick from the load. A contact rated for 230 V that switches a motor sees considerably more across its gap at opening.
Battery systems. A 48 V nominal battery is 58 V at float charge, and the working voltage for clearance purposes is the higher figure.
Establishing it honestly
The working voltage should come from the schematic and from measurement, not from the product label. Measure the node with an instrument that captures peaks, at the worst operating condition — maximum load, minimum input voltage, highest temperature.
Where a transient is expected, the impulse requirement is added to the steady-state working voltage, and the two together define what the insulation must withstand.
What to record
The working voltage at each insulation under assessment, how it was determined (calculated or measured), the nominal supply, and the clause that maps them to a clearance.
A design file that records only the nominal voltage has lost the one number that would justify its clearances, and the next engineer will re-derive them — probably differently.