IEC 61010
 
IEC 61010

Company News

Pollution Degree: Why the Same Clearance Passes at 2 and Fails at 3

Two products, same voltage, same gap between conductors, one passes and one fails. The difference is the pollution degree the design assumes, and it is the parameter most often inherited from an earlier design without checking.

What pollution degree describes

It describes the expected contamination of the environment around the insulation. Four levels, from clean to conductive pollution. What changes between them is not the air gap itself but what may bridge it.

At degree 2, only non-conductive pollution occurs, with occasional temporary condensation. At degree 3, conductive pollution occurs, or dry non-conductive pollution becomes conductive through condensation.

DegreeEnvironmentEffect on clearance
1sealed, climate controlledsmallest required
2office, laboratorymoderate
3workshop, industriallarger, often 1.5 times or more

Why one step changes so much

The tables in IEC 61010 give a clearance for a voltage and a degree. Going from degree 2 to 3 typically adds a factor of 1.3 to 1.6 for the same voltage, because the design must now tolerate a conductive bridge that forms and later dries.

The mechanism is specific. Conductive dust settles on the surface. It is dry and harmless until humidity rises; then it becomes a resistive path. That path passes a small current, which heats the surface, which drives off moisture, which interrupts the path, which lets it cool and re-condense. The cycle repeats, carbonises the surface, and eventually a permanent conductive track forms.

A clearance adequate for degree 2 is not adequate for this, because the failure is not a spark across air. It is a surface process that the air gap cannot interrupt.

Where the assumption goes wrong

Three ways a degree 2 design ends up in a degree 3 environment:

Copying an earlier design. The predecessor was a lab instrument; the successor is sold for workshop use, and the clearance table was not revisited because the voltage had not changed.

Ignoring the customer. The end user installs the unit next to a grinding machine. The documentation says degree 2, but nothing in the product prevents that installation.

Coating applied without redesign. A conformal coating can raise the effective pollution degree, but only if it covers the entire creepage path continuously and is qualified for the environment. A coating with a pinhole is a degree 3 surface with a degree 2 table.

Creepage versus clearance

Both are affected, but differently. Clearance is the air gap; pollution degree affects it because polluted air and surface films reduce the effective withstand. Creepage is the surface path; it is affected more directly, because the surface is where the conductive film forms.

In practice, when a design moves from degree 2 to 3, the creepage distance grows more than the clearance. Designs that address only clearance still fail.

Establishing the degree honestly

The degree is a property of the installation, not of the product alone. A product can be designed for degree 2 and marked accordingly, with instructions that it be installed in a controlled environment. That is legitimate, provided the marking and the instructions actually reach the installer.

What is not legitimate is designing to degree 2 and assuming the customer will provide degree 2 conditions without being told. That assumption is the origin of most field failures in this category.

What to record

The pollution degree assumed, the clause it came from, the resulting clearance and creepage, and the installation conditions the assumption depends on. Four items, and the fourth is what turns a design decision into a defensible one.