The protective bonding test is one of the simplest in the standards: pass a defined current between the protective terminal and an accessible metal part, measure the voltage drop, calculate the resistance. Pass if it is below the limit. It is also one of the most over-interpreted.
What the test measures
The measurement is a four-wire resistance at a specified current — commonly 25 A for equipment, less for some product categories. The current is chosen to be well above any measurement current, so that contact resistance at the joints is measured under something closer to service conditions.
That is the point of using 25 A rather than a milliohmmeter at 200 mA: a joint with a small contact area reads differently at high current, because the contact points heat and soften. A joint that measures 0.3 ohm at 200 mA may measure 0.08 ohm at 25 A.
| Measured at | Good joint | Marginal joint |
|---|---|---|
| 200 mA | 0.05 ohm | 0.08 ohm |
| 25 A | 0.07 ohm | 0.60 ohm |
What it does not prove
Three things, all of which matter in service:
Behaviour at fault current. A 25 A measurement is a fraction of the current a short circuit will drive. A joint with a single point contact that survives 25 A can weld open or vaporise at 1000 A. The test does not extrapolate.
Behaviour over time. Corrosion, vibration loosening and thermal cycling all change a joint after the test. A pass at the factory says nothing about year five.
Behaviour at the other joints. The test measures the path from one accessible part to the terminal. A product with several accessible parts needs each one tested, because the paths differ.
The joints that fail in the field
Three patterns account for most failures:
Paint under the terminal. The screw contacts paint rather than metal. Resistance measures high, or measures low because the screw has cut through the paint at one point — which then corrodes.
Wrong torque. Under-torque gives a small contact area; over-torque damages the thread or the terminal, and the joint loosens later.
Dissimilar metals. A steel screw into an aluminium casting, in a humid environment, corrodes at the interface. The resistance rises over months, and the product that passed at the factory fails at the next inspection.
How to measure it properly
Four-wire, always. A two-wire measurement includes the lead resistance, which at 25 A through a one-metre lead can be tens of milliohms — comparable to the limit itself.
Measure at bare metal. A probe on a painted surface measures the paint. Use a sharp probe that penetrates coating, or find an unpainted feature.
Measure every accessible part, not a sample. The parts with the longest path, or with the most joints in series, are the ones that fail.
What the record needs
Test current, the two points measured, the value, the instrument, and the ambient temperature. Copper resistance rises about 4 % per 10 °C, which is enough to matter when a result is close to the limit.
A record of "bonding OK" without the points is not a record. The whole value of the test is that it locates the failure, and a value without a location locates nothing.