Two currents flow in any earthed piece of equipment: one through the user if they touch the enclosure, and one down the protective conductor to earth. Both are specified, both are limited, and they are not the same measurement even on the same product.
The difference in one sentence
Touch current is what flows through a person touching an accessible part. Protective conductor current is what flows through the earth conductor when nobody is touching anything.
The first is a hazard to a person. The second is a hazard to the installation — it can trip RCDs, overload shared earth paths, and indicate insulation degradation. A product can be safe to touch and still have an excessive earth current.
| Measurement | Circuit | Limit depends on |
|---|---|---|
| touch current | network to earth, person in the path | product class, environment |
| protective conductor current | earth conductor directly | conductor size, installation |
Why the numbers differ so much
On a Class I product with a filter, the Y-capacitors connect live to earth. At 50 Hz, their impedance is low enough that a few milliamps flow continuously down the protective conductor. That current is the protective conductor current, and it can be several times the touch current limit.
The touch current is lower because the person does not provide a perfect path — the measurement network includes a resistance that limits the current. But that resistance is what makes touch current a measure of hazard rather than of leakage.
This is why a product can have 3.5 mA of earth leakage and still pass the touch current requirement: the leakage is real, but the path through a person is limited by the network.
What changes the readings
Configuration. Touch current is measured with the product in normal operation and with each accessible part contacted in turn. Protective conductor current is measured with the product earthed normally and nothing touched.
Voltage. Both are measured at the specified supply voltage, which is usually above nominal. A measurement at nominal understates both.
Frequency. Filter capacitors behave differently at 50 and 60 Hz, and at the higher frequencies produced by switching supplies. The measurement network is specified to weight the frequencies that matter for a shock hazard, so the two measurements respond differently to the same harmonic content.
Load. A switch-mode supply draws different leakage at different loads, because the switching frequency and duty cycle change. Light-load measurements are commonly the lowest and the least representative.
The failure that only one test catches
A degraded Y-capacitor increases earth leakage without changing the touch current much, because the capacitor is between live and earth rather than between live and the accessible surface. Only the protective conductor current sees it.
Conversely, a breakdown in basic insulation between a live part and an accessible metal part increases touch current sharply while earth leakage may rise only slightly, because the current now has a path through the person rather than through the earth conductor.
Measuring only one of the two leaves one of these failures invisible.
Practical setup notes
The measurement network must be the one the standard specifies — using a generic leakage clamp gives a number that cannot be compared with a limit written for a weighted network.
The earth connection during the touch current test is important. If the protective conductor is left connected while the network is inserted, part of the current bypasses the measurement and the reading is low.
Record the supply voltage, the load condition, which accessible parts were contacted, and the network used. Those four items are what makes the result reproducible by someone else.