Protective earth continuity is the test everyone agrees is important and nobody enjoys running, because it needs a current source that most benches do not have permanently connected. The result is that it gets done once at type test and then assumed.
Why the test current is high
A bond is not a resistor. It is a joint — a screw into a chassis, a crimp on a wire, a welded stud. Joints have contact resistance that is non-linear with current: at 200 mA a slightly oxidised joint may read 0.08 ohm; at 25 A the same joint can read 0.6 ohm because the current path is forced through a smaller area of true contact.
That is the reason the standards specify a high test current. The bond has to carry fault current long enough for the protective device to operate, and the resistance at that current is what matters — not the resistance at a multimeter's test current.
| Test current | Typical reading on a good joint | On a marginal joint |
|---|---|---|
| 200 mA (multimeter) | 0.05 ohm | 0.08 ohm |
| 10 A | 0.06 ohm | 0.25 ohm |
| 25 A | 0.07 ohm | 0.6 ohm |
Where the resistance actually sits
On a typical instrument, the bond path has four joints and the resistance is rarely where people look:
The protective conductor terminal. Often the worst joint, because it is assembled once and never re-torqued. Paint under the screw head is a classic.
The chassis connection. A screw into a painted panel relies on the thread cutting through paint, which works when new and degrades with thermal cycling.
The internal wiring crimp. A correct crimp is the lowest-resistance element. A crimp made with the wrong die is the highest.
The measurement point. Probing on a painted surface instead of bare metal adds contact resistance that belongs to the probe, not the product.
Test setup that gives repeatable numbers
The current is passed between the protective conductor terminal and a point on the earthed metalwork, and the voltage drop is measured across the same two points with a separate pair of leads. Using the current leads for the voltage measurement adds the lead resistance to the result, which on a 25 A test can be tens of milliohms.
The measurement point must be bare metal. A file or a sharp probe through paint is acceptable for the test; the point should be chosen where the operator can reach it and where the mark will not matter.
What the result means over time
A new instrument typically measures 0.05 to 0.1 ohm. The value creeping upward over successive annual tests is the useful signal: 0.08 to 0.12 to 0.2 says the joints are degrading, long before the limit is reached.
Recording the number, not just pass or fail, turns the annual test into a trend. That is the difference between a test that satisfies an auditor and a test that prevents a failure.