contact us
What Is Automatic Test Equipment (ATE)?
products
- Main Products
technical articles
Company News
Earth continuity is one of the easiest tests to perform and one of the easiest to perform wrongly. A multimeter and two probes will give a reading, the reading will be in milliohms, and the product will pass. And then the joint will overheat under a fault current because it was never asked to carry one.
Why the current matters as much as the resistance
Resistance is not a constant. A joint that measures 0.05 Ω at 200 mA can measure 0.5 Ω at 25 A if the contact is a single point, a screw thread with paint in it, or a crimp that has not fully collapsed. The measurement current has to be large enough to break down the thin films that sit between two metal surfaces, which is exactly what a fault current would do.
The test current in most safety standards is 25 A for a short duration, or 1.5 times the rating of the circuit, whichever is greater. That current is not there to heat the joint; it is there to make the measurement honest.
| Measurement method | Current | What it proves | Risk if used alone |
| Multimeter continuity | 1 to 10 mA | A path exists | Passes high-resistance joints |
| Micro-ohmmeter, 4-wire | 100 mA to 1 A | Resistance accurately at low current | Still misses current-dependent joints |
| Bonding tester at 25 A | 25 A, 1 to 10 s | Resistance under fault current | Needs a proper tester and leads |
| Visual plus torque check | None | Mechanical integrity | Cannot quantify |
| Voltage drop at rated current | Rated | Real operating condition | Slower, needs a load |
The four joints that fail
Paint or anodising under a bonding screw. The screw cuts through some of it and the rest acts as a resistive layer that breaks down only at high current. Star washers help, and removing the coating at the contact area helps more.
A crimp on a conductor with the wrong die. The crimp looks sound and holds the wire, but the contact area is a fraction of what it should be. A pull test passes, a resistance test at low current passes, and a 25 A test fails.
A bonding stud fixed into a plastic boss with a metal insert. The insert is bonded to the chassis through a thread that was never designed to carry current. It measures fine until it heats.
Two painted surfaces bolted together with no masking. The paint is an insulator; the bolted area is the only conductor, and the contact pressure is a function of a torque that nobody checked.
Where to put the probe
Measure between the protective conductor terminal and the part that a person can touch, not between two points on the chassis. A chassis with a good bond to the terminal and a bolted panel with a bad bond to the chassis has two measurements, and only the second one matters for safety.
- Probe on bare metal, never through paint, at both ends.
- Apply the test current for the specified duration and read the value at the end, not at the start.
- Record the value in milliohms; a pass at 0.09 Ω against a 0.1 Ω limit is a warning, not a comfort.
- Repeat after the vibration or impact test; a joint that passes cold can fail after mechanical stress.
The limit itself
The 0.1 Ω figure that appears in many product standards is not arbitrary: it is roughly the resistance at which a fault current of 25 A develops less than 62 W in the joint, which is survivable for the short time before the protective device operates. A joint at 0.5 Ω develops 312 W in the same event, and that is a fire risk in a plastic enclosure.
Standards referenced: IEC 61010-1 clause 6.5 (protective bonding), IEC 60335-1 clause 27 (earthing), IEC 62368-1 clause 5.6.4 for the equivalent requirement in audio/video equipment.





