The cord anchorage test is a pull and a twist, followed by a measurement. The pull is easy to pass. The measurement is where products fail, because a cord can be held firmly at the gland and still transmit movement to the terminals inside.
What the test does
The cord is pulled with a specified force, usually 30 to 60 N depending on mass, for 25 cycles of one second. Then it is twisted through a specified angle, usually 180 degrees, for a number of cycles. After that, the conductors must not have moved more than 2 mm in the terminals, and the cord must not be damaged.
| Item | Typical requirement | What fails |
| Pull force | 30 to 60 N, 25 cycles | Gland slips, cord sheath stretches |
| Twist | 180°, 10 cycles | Terminal screws loosen |
| Conductor movement | ≤ 2 mm | Insulation pulled into the terminal |
| Sheath damage | No visible damage | Sharp edge at the gland entry |
| Clearance after test | Creepage and clearance maintained | Cord pulled against a live part |
Why the 2 mm is the real test
A clamp that grips the sheath but leaves 200 mm of free conductor inside the enclosure allows that conductor to act as a spring. Under repeated pull, the strands creep within the terminal, insulation moves toward the clamp, and eventually the terminal is gripping insulation instead of copper.
The failure does not appear during the test. It appears months later, as a hot terminal or an intermittent fault.
Design details that pass
- Clamp the sheath, not the conductors, and leave the clamp close to the terminals.
- Use a cable gland with a defined grip range; an oversized gland slips.
- Route the cord so the pull is along the gland axis, not across it.
- Terminate with a ferrule or a crimped sleeve so strands cannot creep.
Standards referenced: IEC 61010-1 clause 6.10 and Annex K, IEC 60320 for the appliance coupler if one is used.
The cord type matters
A cord with a tough outer sheath resists pull better than a thin-walled one, but the sheath also transmits less grip to the gland. The most common failure is a sheath that is too smooth: the gland clamps, the cord creeps, and 0.5 mm of movement per cycle adds up.
Where the cord is smooth, a gland with a serrated insert or a separate clamp plate gives a defined grip.
Testing after assembly, not before
The anchorage test is done on the assembled product, with the cord in its final routing. Testing a bare gland on a bench proves the gland works, not that the product does.
What 2 mm of movement looks like
Mark the conductor at the terminal with a fine line before the test and measure the displacement afterwards. A movement of 2 mm is visible; 0.5 mm is not, and it is the 0.5 mm that accumulates over a thousand handling cycles in the field.
Record the measurement, not just pass or fail. The number tells you how close the design is to the limit, which is what a follow-up change needs to know.
The order of the two tests
Pull first, then twist, then measure. Twisting after pulling tests whether the pull loosened the terminal; twisting first can mask a marginal anchorage by redistributing the conductor.