A gap is not a distance until you know what surrounds it. IEC 61010-1 builds its insulation requirements on two environmental factors, pollution degree and material group, and then reads two limits out of a table. The same physical gap can pass or fail depending on those two inputs.
What pollution degree describes
Pollution degree 1 is a sealed, dry environment. Degree 2 is a normal laboratory or office. Degree 3 is an industrial environment with conductive dust or condensation. Degree 4 is outdoor or continuous conductive contamination.
Most laboratory equipment is degree 2. Equipment intended for a workshop or a plant room is degree 3, and the creepage distances roughly double.
| Pollution degree | Typical environment | Creepage at 250 V, material group IIIa | Note |
| 1 | Sealed, dry | 1.0 mm | Rarely claimed without a sealed enclosure |
| 2 | Lab, office | 2.5 mm | Default for indoor instruments |
| 3 | Industrial, dust or condensation | 4.0 mm | Requires a design decision |
| 4 | Outdoor, conductive contamination | Not permitted without special measures | Usually avoided |
Material group, and why CTI matters
Material group I has a comparative tracking index of 600 V or more, group II is 400 to 599, group IIIa is 175 to 399. A higher CTI allows a shorter creepage distance for the same pollution degree, which is why flame-retardant filled plastics are common in compact equipment.
The design sequence that avoids rework
- Decide the pollution degree from the intended environment, not from the prototype.
- Get the CTI of the actual moulding compound, not the family datasheet.
- Read clearance and creepage from the table for the working voltage.
- Add the tolerance of the moulding process to the measured gap, not to the nominal.
The last point is the one that fails in production: a design with 2.5 mm nominal creepage and a ±0.3 mm moulding tolerance can measure 2.2 mm, which is below the limit.
Standards referenced: IEC 61010-1 clause 6.7 and Annex K, IEC 60664-1 for insulation coordination, IEC 60112 for CTI measurement.
Clearance and creepage are not the same measurement
Clearance is the shortest distance through air. Creepage is the shortest distance along a solid surface. A slot in the housing increases creepage without changing clearance, which is why slots are used to add insulation distance in a compact design.
Measure them separately: a rule for clearance, a fine wire or a profile projector for creepage. Using one number for both is a common cause of a failed design review.
The working voltage, not the mains voltage
The tables are indexed by working voltage, which for a switched-mode supply may be 400 V DC on the primary side even though the input is 230 V AC. Using 230 V in the table gives a distance that is too short.
Coating is not a substitute
A conformal coating can reduce the required creepage under some standards, but only if it is qualified as a type 2 coating and applied over a defined area. A spray that covers most of the board is not the same thing, and auditors know it.
Where the design is tight, a slot or a rib is more predictable than a coating, and easier to verify in production.