Most of the cost of a safety design is in the separation between the mains side and the accessible side. That separation can be reduced when the accessible side is fed from a limited-energy source, which is why the definition matters commercially as well as technically.
The classical numbers are 30 V and 240 VA, and they come from the idea that a source below them cannot deliver a dangerous shock or start a fire through a small fault.
Where the numbers come from
| Parameter | Limit | Reason |
| Voltage | 30 V AC or 60 V DC | Below the threshold for dangerous current through dry skin |
| Power | 240 VA for 1 min | Cannot ignite common materials through a small fault |
| Current into a short | 8 A for 1 min at the rated voltage | Fire initiation threshold in practice |
| Energy storage | Limited by the same 240 VA rule | A capacitor bank can deliver more than its supply |
The power limit is measured, not calculated: short the output through the specified load and watch the source for a minute. A supply that limits at 30 V but delivers 400 VA into a short is not compliant however comfortable 30 V feels.
What the classification buys
- Reduced spacing. Clearance and creepage between a limited-energy circuit and an accessible part can be smaller than between mains and accessible.
- Simpler fault analysis. Shorting a limited-energy node cannot produce a fire, so the component-level analysis narrows.
- Different marking. Some standards allow a limited-energy output to be treated as an output terminal rather than a hazardous live part.
Three ways designs lose the classification
First, a capacitor on the output stores enough energy to exceed the limit for a fraction of a second, and the rule looks at the energy, not only the continuous power. Second, a bridging fault inside the supply connects the mains side to the output through a single component, which is exactly the single-fault case. Third, the measurement is taken at 25 °C while the supply is rated to 40 °C and limits less power when hot.
Our electrical safety test equipment includes the short-circuit load bank and the energy measurement module, and the technical notes give the test sequence for the classification.
Standards referenced: IEC 61010-1 clause 6.3.2 and 6.4, IEC 60950-1 clause 2.5 (limited power source), IEC 62368-1 clause 6.2 for the newer PS classification.
The measurement that settles the argument
Short the output through a 0.1 Ω resistor and record voltage and current for 60 s. Multiply them at each reading and take the maximum: above 240 VA for more than a second means the circuit is not limited, whatever the open-circuit voltage says.
Repeat at 40 °C: a supply that limits at 25 °C may deliver 15 percent more current when cold and 10 percent less when hot, and only the hot case is the declared one.